MESSAGE PROBLEM WHAT TO DO
4) There was not enough Repeat the test with a new
blood on the Test Strip. Test Strip and a larger drop
of blood.
5) The Meter was used in Move the Meter out of
very bright light. direct light and repeat the
test with a new Test Strip.
6) The Check Strip procedure Repeat the Check Strip test.
was incorrect.
7) The Meter may not be Call LifeScan Technical
operating correctly. Services.